Observe the efficacy of pacemaker in transurethral bipolar vaporization of prostate
-
摘要: 目的:观察前列腺双极等离子汽化电切术对心脏起搏器工作的影响。方法:选取22例携带有心脏起搏器的前列腺增生患者行前列腺双极等离子气化电切术,术前行24 h动态心电图检查,术中密切监护心电变化,术后24 h心电监护,术后1周行24 h动态心电图检查以了解起搏器的工作状态的变化。结果:术前术后起搏器工作状态未发生改变,术中心电图也未受到明显干扰,手术顺利完成,围手术期心功能稳定。结论:未发现前列腺双极等离子气化电切术对心脏起搏器的工作产生干扰,留置心脏起搏器并非是前列腺等离子双极汽化电切术的禁忌证。
-
关键词:
- 前列腺双极等离子汽化电切 /
- 起搏器 /
- 前列腺增生
Abstract: Objective:Objective to observe the efficacy of pacemaker in transurethral bipolar vaporization of prostate.Method:Twenty-two patients of benign prostate hyperplasia with permanent pacemaker implantation were selected,To understand the working status of the change of the pacemaker rely on the results of the Dynamic Electro-cardiogram in preoperative. The electrocardiographic monitor in operation and after surgery 1 week review Dynamic Electro-cardiogram.Result:Pacemakers working state not changed before the operation and after surgery,Intraoperative electrocardiogram were not obvious interference, Operation successfully completed,Perioperative heart function stability.Conclusion:No interference to pacemaker was found in transurethral bipolar vaporization of prostate. The patients with permanent pacemaker implantation is not the contraindication of transurethral bipolar vaporization of prostate. -
-
[1] 彭小红,张小岗.动态心电图在起搏器功能异常检测中的价值[J].现代医药卫生,2008,22:3392-3393.
[2] SWEESY M W,HOLLAND J L,SMITH K W.E-lectromagnetic interference in cardiac rhythm manage-ment devices[J].AACN Clin Issues,2004,15:391-403.
[3] 庄剑秋,郎根强,张炯.前列腺增生并置心脏起搏器经尿道前列腺汽化切除术6例报告[J].现代泌尿外科杂志,2009,14:295-297.
[4] 季惠翔,张家华,宋波,等.安置心脏永久性起搏器患者的经尿道前列腺电切治疗[J].中华男科学杂志,2005,11:530-535.
[5] 那彦群.吴阶平泌尿外科学[M].济南:山东科学技术出版社,2009:1183.
-
计量
- 文章访问数: 79
- PDF下载数: 78
- 施引文献: 0